\(\renewcommand\AA{\unicode{x212B}}\)
CombineDiffCal v1¶
Summary¶
Combine a per-pixel calibration with a grouped spectrum calibration
Properties¶
| Name | Direction | Type | Default | Description | 
|---|---|---|---|---|
| PixelCalibration | Input | Mandatory | DiffCal TableWorkspace that will be updated. This is often generated from cross-correlation. These are the “prev” values in the documentation. | |
| GroupedCalibration | Input | Mandatory | DiffCal table generated from calibrating grouped spectra. This is the “DIFCpd” value in the documentation. | |
| CalibrationWorkspace | Input | Mandatory | Workspace where conversion from d-spacing to time-of-flight for each spectrum is determined from. This is the “DIFCarb” value in the documentation. | |
| OutputWorkspace | Output | DiffCal table generated from calibrating grouped spectra | ||
| MaskWorkspace | Input | MaskWorkspace | MaskedWorkspace for PixelCalibration | 
Description¶
This algorithm combines pixel calibration results (\(prev\) subscript below) from the PixelCalibration workspace with the calibration of focussed data from PDCalibration (\(pd\) subscript below).
The \(arb\) subscript below denotes values found in the CalibrationWorkspace.
This is part of the larger workflow for powder diffraction calibration.
Warning
The GroupedCalibration table is assumed to not have \(DIFA\) or \(TZERO\) set.
Input arguments¶
- PixelCalibrationis the previous calibration result- Has values for each detector pixel 
- Provides \(DIFC_{prev}\), \(DIFA_{prev}\), and \(TZERO_{prev}\) 
 
- GroupedCalibration- Has values for each detector pixel, but the values are generally the same within a focused group 
- Provides \(DIFC_{pd}\) 
- Is the - OutputCalibrationTablefrom previous PDCalibration execution
 
- CalibrationWorkspace- Has values for each spectrum which are grouped pixels 
- Provides \(DIFC_{arb}\) 
- Is the - InputWorkspacefrom previous PDCalibration execution
 
How values are combined¶
The effective calibration values are calculated using the following equations
where \(DIFC_{pd}\)  is from PDCalibration,  \(DIFC_{arb}\) is found in the parameter CalibrationWorkspace that was the input workspace to PDCalibration, and \(DIFC_{prev}\) is from the previous calibration.
The value of \(TZERO\) is unchanged
The value of \(DIFA\) is updated by
If pixels are masked in MaskWorkspace, missing from PixelCalibration, or missing from the list of contributing pixels in CalibrationWorkspace, then the calibration constants found in GroupedCalibration will be copied.
Categories: AlgorithmIndex | Diffraction\Utility
Source¶
C++ header: CombineDiffCal.h
C++ source: CombineDiffCal.cpp