Texture Experiment Planner#
Interface Overview#
This custom interface integrates the tasks involved in planning and executing texture experiments (see Texture Analysis Theory and Texture Reduction for more context). It is primarily intended to serve ENGINX and IMAT; however, it is compatible with all Mantid instruments.
The general workflow of the interface is:
Define an experimental sample with the correct shape, material properties, and initial position on the beamline.
Define the sample directions (see Texture Analysis Theory for more details).
Set the instrument and virtual detector grouping (this can be updated throughout the process).
Optionally define a gauge volume (otherwise the whole sample is assumed to be illuminated).
Either load or manually add the experimental orientations.
Observe the pole figure coverage that your current experimental setup will yield.
Iterate on the orientations until the desired coverage is achieved.
Optionally, view and optimise the experiment based on the estimated transmission values for each run and each virtual detector.
Export the desired information. This can be the experimental orientations in a variety of formats, the definition of the sample on the beamline to be used as a reference, or the relative weightings of the estimated transmission values.
General Information#
- Red Stars
Red stars next to browse boxes and other fields indicate that the file could not be found. Hover over the star to see more information.
- Settings
The cog icon opens an additional settings menu, which exposes some of the interface behaviour that might otherwise clutter the main window. See Settings Menu for the full set of options.
Sample Setup Options#
- Load Sample Shape
Options to load a shape which has been defined either as a shape mesh (
.stlfile) or using Constructive Solid Geometry (CSG -.xmlfile). See How To Define Geometric Shape for more information on how to format a CSG shape definition.- Set Sample Material
Set the material of the sample using the SetSampleMaterial v1 algorithm.
- Update Initial Shape Orientation
Adjust the starting orientation of the shape relative to the components of the selected instrument. This should match how the sample will be (or is) loaded onto the beamline, before the sample positioner has repositioned it for a given orientation.
- Update Initial Shape Position
Adjust the starting position of the shape relative to the components of the selected instrument. This should match how the sample will be (or is) loaded onto the beamline, before the sample positioner has repositioned it for a given orientation.
- Update Sample Directions
Set the three intrinsic sample directions, relative to which all of the texture information is calculated. The pole figure projection is always taken into the plane of the first and the last directions.
Note that these are defined in terms of the instrument reference frame, not the sample shape that has been loaded and potentially repositioned.
Update Texture Directionsmust be clicked to apply the changes.
Experimental Setup#
- Instrument
Select from the fully supported instruments (ENGINX or IMAT) or try a different Mantid instrument (
Custom). IfCustomis selected,Groupis fixed toCustomas well and the relevant dialog boxes will appear.- Group
Select the desired virtual detector grouping. Some pre-set options are available for both ENGINX and IMAT, but custom XML definitions can be loaded in addition.
- Custom Instrument
Type the name of the Mantid instrument you would like to use for the experiment. The interface checks the input text against the instrument definitions it can find. If the text does not match a known instrument the box will appear red and
Update Instrumentwill remain disabled.- Grouping File
Select a detector grouping file (see SaveDetectorsGrouping v1 and LoadDetectorsGroupingFile v1) to be used. If the file is incompatible with the selected instrument,
Update Instrumentwill remain disabled.- Update Instrument
Update the experiment with the current set of instrument options.
- Gauge Volume Options
Set an experimental gauge volume either from the pre-set definitions or from a custom CSG definition (see How To Define Geometric Shape).
- Set Gauge Volume
Apply the current gauge volume definition to the experiment.
- Clear Gauge Volume
Remove the gauge volume definition - a shortcut for setting
No Gauge Volume.- Load Orientation File
Load a set of orientations from a
.txtfile of either Euler angles or flattened rotation matrices (see the orientation file section of the Texture Analysis Concept doc). The file type is automatically determined from the number of entries in each row.The axes and senses of the Euler angles can be set in the settings menu, and these axes also define how rotation matrices will be displayed as decomposed axes.
Once loaded, these orientations populate the Orientation Table.
Goniometers#
- Num axes
Number of active goniometer axes for the sample positioner.
- Step size
Number of degrees that a single arrow click for the
Anglefields will increment or decrement.- Current Orientation Index
The row of the Orientation Table currently being shown in the Axes below and highlighted in the Pole Figure Display (when not in the transmission estimation view).
- Add Orientation
Add a duplicate of the current orientation to the end of the Orientation Table.
- Axes
Display a
Vector,SenseandAnglefor up to 6 dependent goniometer axes. Axis 0 is the outermost, and each subsequent rotation is intrinsically linked to the rotations of the outer axes. The goniometers can be visualised by enablingShow Goniometersin the settings.
Lab View#
- Figure
Displays the sample oriented on the chosen instrument. From the settings it is possible to toggle the visibility of:
The sample direction vectors
The goniometers
The incident neutron beam
The diffraction vectors (Ks)
The scattered detector vectors
Pole Figure Display#
- Figure
Shows the coverage for the current set of orientations with
Includeselected in the Orientation Table.The
Current Orientation Indexfrom theGoniometerspanel is displayed as solid blue points (if included, otherwise as grey point outlines), and the other orientations as blue point outlines.The goniometer axes are displayed as coloured points which match their colour in the Lab View display (the goniometers can be visualised by enabling
Show Goniometersin the settings). Goniometer axes which lie within the plane of the projection are displayed as lines. The most recently updated goniometer is displayed as a solid point or line, with the others as outlines or dashed lines.- Estimate Transmission Values
When enabled, the figure instead displays the estimated transmission values for each virtual detector, based on a Monte Carlo simulation at a specified TOF/d-spacing/wavelength (these can be modified in the settings menu).
Orientation Table#
- Orientations
Contains the axis information for all of the experimental orientations, as well as which ones should be included in the Pole Figure Display and any outputs (the
Includetoggle). Additionally there is aSelecttoggle which allows several orientations to be removed at once.- Select All / Deselect All
Tick or clear the
Selectbox on every orientation in the table.- Delete Selected
Remove all orientations whose
Selectbox is ticked.
Output Options#
- Save Directory
The directory to save the output files into.
- Filename
The file name for the output file.
- Export
The type of file to write (see the options below).
- SScanSS-2 angles
Write the included orientations into a
.anglesfile which can be loaded into SScanSS-2.- Euler Orientation file
Write the included orientations into an Euler angle
.txtfile. The axes used are the axes defined in the settings.- Matrix Orientation file
Write the included orientations into a flattened rotation matrix
.txtfile.- Reference Workspace
Write the defined sample onto a reference workspace
.nxsfile. This can be loaded directly into the Absorption Correction tab of the Engineering Diffraction interface to start processing results.- Transmission Weighting
Write a
.txtfile with one line per included orientation. For each line the value given is the transmission value of the most absorbing virtual detector, normalised against the orientation with the highest of these values. This gives an estimate of the relative counting time required to match the scattered intensity of the least absorbing orientation.